NSN 6530-01-431-5146
Part Details | PORTABLE VOLUME VENTILATOR
6530-01-431-5146 An item designed to provide ventilation to a patient in transport and/or during institutional care. It may be in a transport case and provide time cycled, pressure limited, piston driven constant flow, be microprocessor controlled, with three modes, control, assist-control, synchronous intermittent mandatory ventilation providing sensitivity control, rechargeable battery, hoses, tubing, connectors, patient valve, and other related features.
Alternate Parts: 960337, 96-0337, 6530-01-431-5146, 01-431-5146, 6530014315146, 014315146
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 65 | MAY 14, 1996 | 01-431-5146 | 48820 ( VENTILATOR, VOLUME, PORTABLE ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6530-01-431-5146
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| 96-0337 | 89875 | DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL |
Technical Data | NSN 6530-01-431-5146
| Characteristic | Specifications |
|---|---|
| CURRENT TYPE | B$C AC/DC |
| PORTABILITY | PORTABLE |
| OVERALL LENGTH | 18.000 INCHES NOMINAL |
| OVERALL WIDTH | 18.000 INCHES NOMINAL |
| OVERALL HEIGHT | 21.000 INCHES NOMINAL |
| FURNISHED ITEMS AND QUANTITY | OPERATORS, 2; SERVICE MANUALS, 2 |
| SPECIAL FEATURES | THE UNIT IS A TRANSPORT VENTILATOR FOR AIR AND GROUND MOVEMENT OF PATIENTS. IT IS A SELF-CONTAINED SYSTEM CAPABLE OF SUPPLYING 21% TO 100% OXYGEN WHEN PROVIDED 50 PSI OXYGEN SOURCE. THE BREATHING CIRCUIT USES STANDARD DISPOSABLE UNIVERSAL 22 MM TUBING SET. FUNCTION MODES: ASSIST/CONTROL; INTERMITTENT MANDATORY VENTILATION (IMV); CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP); AND POSITIVE END EXPIRATORY PRESSURE (PEEP). ALARMS: APNEA, HIGH PRESSURE, LOW PRESSURE (DISCONNECT), EXTERNAL POWER (LOW/FAIL), BATTERY (LOW/FAIL), LOW GAS SOURCE PRESSURE, LOW PEEP AND LOW EXHALE VOLUME. VISUAL: BREATHS PER MIN., PEAK INSPIRATORY PRESSURE (PIP), MEAN AIRWAY PRESSURE (MAP), PEEP, EXHALE VOLUME, SPONTANEOUS PATIENT RATE, AND ALARM STATUS. THE UNIT ALLOWS VENTILATION OF BOTH CHILDREN (1-12 YEARS) AND ADULTS. ELECTRICAL: 115/230 VAC, 50/60/400 HZ, 11-30 VDC; JOINT AIR CERTIFICATION FOR FIXED & ROTARY WING AIRCRAFT FOR COMBINED ESSECTIAL CHARACTERISTICS |
| TEST DATA DOCUMENT | 96906-1472 STANDARD (INCLUDES INDUSTRY OR ASSOCIATION STANDARDS, INDIVIDUAL MANUFACTUREER STANDARDS, ETC.). AND 80204-ES1-1999 STANDARD (INCLUDES INDUSTRY OR ASSOCIATION STANDARDS, INDIVIDUAL MANUFACTUREER STANDARDS, ETC.). |
| SPECIAL TEST FEATURES | JET AIRCRAFT RANDOM VIBRATON TESTING |
| SUPPLEMENTARY FEATURES | UNIT SHALL BE COMPATIBLE WITH THE FOLLOWING REGULATORS: PRESSURE REGULATOR, OXYGEN, FOR D-CYLINDERS, NSN 6680-01-505-1310 AND PRESSURE REGULATOR, OXYGEN, FOR H-CYLINDERS, NSN 6680-01-505-1303 |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | VENTILATOR VOLUME PTBL: |