Part Details | PORTABLE VOLUME VENTILATOR

6530-01-431-5146 An item designed to provide ventilation to a patient in transport and/or during institutional care. It may be in a transport case and provide time cycled, pressure limited, piston driven constant flow, be microprocessor controlled, with three modes, control, assist-control, synchronous intermittent mandatory ventilation providing sensitivity control, rechargeable battery, hoses, tubing, connectors, patient valve, and other related features.

Alternate Parts: 960337, 96-0337, 6530-01-431-5146, 01-431-5146, 6530014315146, 014315146

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
65MAY 14, 199601-431-514648820 ( VENTILATOR, VOLUME, PORTABLE )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6530-01-431-5146
Part Number Cage Code Manufacturer
96-033789875DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL
Technical Data | NSN 6530-01-431-5146
Characteristic Specifications
CURRENT TYPEB$C AC/DC
PORTABILITY PORTABLE
OVERALL LENGTH18.000 INCHES NOMINAL
OVERALL WIDTH18.000 INCHES NOMINAL
OVERALL HEIGHT21.000 INCHES NOMINAL
FURNISHED ITEMS AND QUANTITYOPERATORS, 2; SERVICE MANUALS, 2
SPECIAL FEATURESTHE UNIT IS A TRANSPORT VENTILATOR FOR AIR AND GROUND MOVEMENT OF PATIENTS. IT IS A SELF-CONTAINED SYSTEM CAPABLE OF SUPPLYING 21% TO 100% OXYGEN WHEN PROVIDED 50 PSI OXYGEN SOURCE. THE BREATHING CIRCUIT USES STANDARD DISPOSABLE UNIVERSAL 22 MM TUBING SET. FUNCTION MODES: ASSIST/CONTROL; INTERMITTENT MANDATORY VENTILATION (IMV); CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP); AND POSITIVE END EXPIRATORY PRESSURE (PEEP). ALARMS: APNEA, HIGH PRESSURE, LOW PRESSURE (DISCONNECT), EXTERNAL POWER (LOW/FAIL), BATTERY (LOW/FAIL), LOW GAS SOURCE PRESSURE, LOW PEEP AND LOW EXHALE VOLUME. VISUAL: BREATHS PER MIN., PEAK INSPIRATORY PRESSURE (PIP), MEAN AIRWAY PRESSURE (MAP), PEEP, EXHALE VOLUME, SPONTANEOUS PATIENT RATE, AND ALARM STATUS. THE UNIT ALLOWS VENTILATION OF BOTH CHILDREN (1-12 YEARS) AND ADULTS. ELECTRICAL: 115/230 VAC, 50/60/400 HZ, 11-30 VDC; JOINT AIR CERTIFICATION FOR FIXED & ROTARY WING AIRCRAFT FOR COMBINED ESSECTIAL CHARACTERISTICS
TEST DATA DOCUMENT96906-1472 STANDARD (INCLUDES INDUSTRY OR ASSOCIATION STANDARDS, INDIVIDUAL MANUFACTUREER STANDARDS, ETC.). AND 80204-ES1-1999 STANDARD (INCLUDES INDUSTRY OR ASSOCIATION STANDARDS, INDIVIDUAL MANUFACTUREER STANDARDS, ETC.).
SPECIAL TEST FEATURESJET AIRCRAFT RANDOM VIBRATON TESTING
SUPPLEMENTARY FEATURESUNIT SHALL BE COMPATIBLE WITH THE FOLLOWING REGULATORS: PRESSURE REGULATOR, OXYGEN, FOR D-CYLINDERS, NSN 6680-01-505-1310 AND PRESSURE REGULATOR, OXYGEN, FOR H-CYLINDERS, NSN 6680-01-505-1303
PART NAME ASSIGNED BY CONTROLLING AGENCYVENTILATOR VOLUME PTBL: