Part Details | FLUXMETER

6625-00-178-9580 An electrical instrument which measures magnetic flux. With appropriate calibration it can be used for measurement of the intensity of a magnetic field. See also METER, FIELD STRENGTH.

Alternate Parts: TS15AAP, 6625-00-178-9580, 00-178-9580, 6625001789580, 001789580

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
66JAN 01, 196100-178-958020901 ( FLUXMETER )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6625-00-178-9580
Part Number Cage Code Manufacturer
TS15AAP80058JOINT ELECTRONICS TYPE DESIGNATIONSYSTEM
Technical Data | NSN 6625-00-178-9580
Characteristic Specifications
FLUX DENSITY1000.0 GAUSS MINIMUM AND 10000.0 GAUSS MAXIMUM
SCALE DIRECTION AND RANGE+12.0 TO +96.0 COUNTERCLOCKWISE
SCALE LINEARITY NONLINEAR
MAGNET GAP RANGE ACCOMMODATED+1.300 TO +1.500 INCHES
MAGNET DIAMETER RANGE ACCOMMODATED+0.000 TO +1.625 INCHES
PROBE HANDLE INCLUDED
MAGNET YOKE QUANTITY2
CASE MATERIAL WOOD
REFERENCE NUMBER DIFFERENTIATING CHARACTERISTICSAS MANUFACTURED BY MARION ELECTRICAL INSTRUMENT CO
FUNCTIONAL CLASSIFICATIONAA-5.6
FUNCTIONAL DESCRIPTIONPORTABLE,PROVIDES QUALITATIVE MEASUREMENTS OF FLUX DENSITIES BETWEEN POLE FACES OF MAGNETS EMPLOYED IN OSCILLATORY CIRCUITS OF X-BAND & S-BAND TRANSMITTERS; CALIBRATED DIRECTLY IN GAUSS UNITS OF FLUX DENSITY; C/O 2 MILLIAMMETERS IN SERIES (A GROUP METER & A PROBE METER),A RHEOSTAT,& ASSOCIATED SHUNTS; PROBE METER IS INSERTED BETWEEN POLES OF MAGNET UNDER TEST & INDICATIONS ARE READ ON GAUSS METER; PROBE METER USES FIELD OF MAGNET UNDER TEST TO PRODUCE A DEFLECTION WHEN A CURRENT IS PASSED THROUGH ITS COIL; CURRENT REQUIRED TO PRODUCE THIS READING IS THEN A MEASURE OF FLUX DENSITY; CONTROLS & INDICATORS ARE LOCATED ON FRONT PANEL
RELATIONSHIP TO SIMILAR EQUIPMENTINTERCHANGEABLE W/ 6625005192407,6625004740880,6625005552939
REFERENCE DATA AND LITERATURETM 11-2559 INSTRUCTION BOOK; AN 16-35TS15-32 MAINTENANCE INSTRUCTION; MFR(S) R/N TS-15A/AP
ENTRY DATE70-04-01