NSN 6630-01-364-8555

Part Details | BLOOD GAS ANALYZER

6630-01-364-8555 An instrument designed for determination of the carbon dioxide combining power, oxygen, carbon monoxide and nitrogen of whole blood or blood plasma. May also be used for micro determinations of gas other than blood gases, developed from solutions, such as organic nitrogen and carbon, urea, amino nitrogen, iodates, sulphates, sugar, calcium lactic acid, potassium, and analysis of gas mixtures.

Alternate Parts: 4300M WPROVISIONING, 4300M W/PROVISIONING, N0059, N0-059, 4300M, GEM STAT, 6630-01-364-8555, 01-364-8555, 6630013648555, 013648555

Supply Group (FSG) NSN Assigned NIIN Item Name Code (INC)
66SEP 26, 199201-364-855538049 ( ANALYZER, BLOOD GAS )
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6630-01-364-8555
Part Number Cage Code Manufacturer
4300M W/PROVISIONING0NTK1DANIEL CONSULTANTS INC
N0-05989875DEFENSE LOGISTICS AGENCYDIRECTORATE OF MEDICAL MATERIEL
4300M0CNN1MALLINCKRODT SENSOR SYSTEMS INC
GEM STAT0CNN1MALLINCKRODT SENSOR SYSTEMS INC
Technical Data | NSN 6630-01-364-8555
Characteristic Specifications
METHOD OF ANALYSIS POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE
TEST TYPE FOR WHICH DESIGNEDBLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT
ANALYSIS TEST/MEASUREMENT RANGEPO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT.
TEST SAMPLE/SPECIMEN SIZE0.5 ML
TEST SAMPLE/SPECIMEN TYPEWHOLE BLOOD
FEATURES PROVIDED DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION
CURRENT TYPE AC
VOLTAGE IN VOLTSB120.0$$B230.0
FREQUENCY RATINGA50.0$A60.0
HEIGHT8.400 INCHES
WIDTH9.100 INCHES
DEPTH18.300 INCHES
SPECIAL FEATURESUTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT.
PART NAME ASSIGNED BY CONTROLLING AGENCYBLOOD GAS ANALYZER 120/230V 50/60 HZ AC: